PATIENT ENTRY Head First
PATIENT POSITION Supine
COIL 8HRBRAIN
PLANE OBLIQUE
SEDESCFLAG 1
SERIES DESCRIPTION AXIAL FLAIR
GRADIENT MODE ZOOM
IMAGING MODE 2D
PULSE SEQUENCE T2flair
IMAGING OPTIONS FC, VBw, Fast, ZIP512
PLUG 25
MONITOR SAR y
FILTCHOICE None
BWRT -1
SATLOCZ2 9990
SATTHICKZ2 80.0
CARDIAC TAG SPACE 7
CARDIAC TAG TYPE None
USERCV3 2.00
USERCV_MASK 8
TE 120.0
# ECHOES 1
TR 8600.0
TI 2150
AUTO TRIG TYPE 0
AUTO TRIG WINDOW 0
FOV 24
SLICE THICKNESS 5.0
SLICE SPACING 1.5
SLOC1 R2.9
SLOC2 A2.6
SLOC3 I31.0
ELOC1 R2.9
ELOC2 A32.5
ELOC3 S102.2
NUMBER OF SLICES 24
MATRIXX (FREQ) 320
MATRIXX (PHASE) 192
SWAPPF A/P
NEX 1.00
CONTRAST No
CONTRAST AM Yes
TABLE DELTA 0.00
RECEIVER BANDWIDTH 19.23
AUTOSHIM Yes
PHASECORR No
AUTO CENTER FREQ Water
FLDIR Slice
PAUSEDELMASKACQ 1
AUTOSUBOPTIONS 0
AUTO SCIC 1
Total# Stations 0
STATION 0
generated on
27-Mar-2009 10:32:05
using   ge2txt   ver 1.3
Written by Perry Horwich, M.D.